Installing the Accela Open Autosampler
Accela Open Autosampler Object Positions
Accela Open Autosampler Object Positions
Defining Object Positions
40
Accela Open Autosampler Hardware Manual
This section contains information about Accela Open Autosampler object positions.
Note Remove the syringe adapter from the injection unit before performing the following
steps.
The objective is to define the reference positions for all Accela Open Autosampler objects.
Make sure to properly mount the tray holders, valve drives, and wash station to the Accela
Open Autosampler x axis. The following describes how to teach the reference position for a
tray holder. The described procedure is common to all Accela Open Autosampler objects.
To teach the reference position for a tray holder
1. Switch on the Accela Open Autosampler power supply.
2. Observe the keypad display.
The model name, Accela Open Autosampler, appears along with the firmware version
number. The Job Queue menu screen then appears.
3. Complete the following sequence (common to all objects):
where the Named Tray Holder represents a predefined tray holder (for example, THldr1
or Stack1).
After selecting Named Tray Holder, the x, y, z positions associated with the object are
displayed
(Figure
33).
Thermo Scientific