Accela Open Autosampler Hardware Manual
To install the power supply
1. Locate the power supply, the DC power cable, and the AC power cable.
2. Set the power supply switch to OFF.
3. Connect one end of the DC power cable to the power supply and the other end to the
POWER connector at the back of the x axis.
4. Connect the female end of the AC power cable to the power supply. Then connect the
male end to an AC power outlet.
To install the LC injection valve
1. Locate the valve drive, which has one clamp that is identical to the other object clamps
(Figure
15).
Figure 15. Installing the injection valve drive
2. Attach the valve drive to the x axis and tighten the mounting screw.
Tip To avoid delay volume between sample injection point and detection, place the
injection valve near the detection device.
3. Connect the control cable from the valve drive to AUX1.
Note For information on installing another type or multiple valve drives, see
"Electrical Connections for 2- or 4-Injection Valves"
The injection valve and rotor are preinstalled on the valve drive.
on
page
37.