Injection Valve Plumbing and Connection to DLW Wash Station
Thermo Scientific
To plumb the injection valve and connect it to DLW Wash Station
1. Connect the injection valve waste line from port 2 (for example, the standard Cheminert
valve type), as shown in
Tip If you use two injection valves in the PAL configuration, connect the second
waste line to the lower front waste port of the DLW Wash Station. If you use only one
connector, apply a dummy plug to the second port on the wash station.
2. Place the waste tube from the DLW Wash Station into the waste container.
IMPORTANT You must position the waste container below the injection valve. Make
sure that the waste liquid flows into the waste container without restriction.
Figure 29. Solvent line plumbing on Cheminert valve types
).
Figure
29, to the front waste port of the wash station.
Accela Open Autosampler Hardware Manual