Instrument Dimensions

Table 2. Accela Open Autosampler specifications (Sheet 3 of 3)
Component
Wash station
Wash station
Instrument Dimensions
8
Accela Open Autosampler Hardware Manual
Specification
• DLW Syringe assembly:
Syringe: Gastight Syringe 100 μL with Valflon tip
Wetted parts: Glass, PTFE and Valflon
Syringe needle: Gauge 22 (18/25 mm OD/0.41 mm ID)
Length: 51 mm, PTFE seal
• DLW Actuator/solenoid: 2/2-way flipper solenoid valve
Wetted parts:
Body: PEEK
Seal material: FFKM (Simriz™)
• DLW Manifold Solvent Selector module and Syringe Adapter assembly:
Wetted parts:
Block: sst 1.4435
Perfluor (O-ring)
• Holding Loop:
FEP Tubing OD 1/16 in., ID 0.50 mm,
length 55 cm Loop Volume: 108 μL
• DLW Flow Diverter
Tube dimensions:
Gauge: 22 (OD 18/25 mm/ID 0.41 mm)
Length: 23 mm
Wetted part: Stainless chromium steel
• DLW Needle Holder assembly
Wetted part: sst 1.4435
• Standard Configuration
Accela 1 (1250 bar), Accela 2 (1250 bar with 6-drawer cooled stack), Accela 3
(QuickQuan™), Accela 4 (EQuan™)
The dimensions of the Accela Open Autosampler are as follows:
• Height: 1045 mm (41.1 in.)
• Depth: 717.5 mm (28.3 in.)
• Width: 676 mm (26.6 in.)
• Weight: 38.5 kg (84.9 lb)