Thermo Scientific Accela Open Autosampler Hardware Manual page 41

Table of Contents

Advertisement

Mounting the DLW Pump Holder
Thermo Scientific
You must always mount the pump holder on the far left of the x axis. The solvent lines are
connected to the holder. Moving across the pump holder with the z axis could kink the
stabilizing wire. The best way to install the holder is by dismounting the drive from the Open
Autosampler x axis.
To mount the pump holder
1. Locate the pump holder and the two supplied Torx screws.
2. Attach the holder to the x axis as shown in
Figure 16. Attaching pump holder bracket to the x axis
The solvent line assembly is not preinstalled at the factory. The correct mounting is
shown in
Figure
17.
3. Using the end of the guide wire with the longer tubing ends, connect the guide wire with
the screw (M4 x 8) and the corresponding serrated lock washer to the back of the module.
Pay attention to the guide groove, which orients the wire
Tip Always install the DLW tubing (from the kit) at the pump module first. Before
you connect the other end of the tubing to the DLW Syringe assembly, make sure that
the assembly is not installed in the injection unit.
2
Installing the Accela Open Autosampler
Installing the Autosampler and DLW
Figure
16.
(Figure
17).
Accela Open Autosampler Hardware Manual