Thermo Scientific Accela Open Autosampler Hardware Manual page 62

Table of Contents

Advertisement

2
Installing the Accela Open Autosampler
Accela Open Autosampler Object Positions
44
Accela Open Autosampler Hardware Manual
Figure 36. Tray type VT54
Correction for Inclination in X, Y, or Z Axes
The following describes the correction using the path to the Utilities section.
To correct for inclination in x, y, or z axes
1. After selecting the Utilities function, Tray, verify whether the tray type assigned
corresponds to the tray placed into the tray holder. If so, check the vial positions at the
three corners of the tray. The F3 function key activates this.
A message box opens, displaying a note: Adjust offset X-, Y-, Z for Position 001.
2. Select OK to adjust all three axes perfectly by the top of the vial. This step is only valid for
position number 001. See
Figure 37. Message to adjust offset x, y, z for position 001
Figure
37.
Thermo Scientific