Differences Between DLW and DLW-2
Thermo Scientific
Table 1
shows the differences between the critical parts or parameters of the DLW and
DLW-2 options.
Table 1. Differences Between DLW and DLW-2
Part or parameter
Holding Loop
ID
Length
Material
OD
Volume
Syringe Needle
Length
Gauge
Material
DLW Flow Diverter
Length
OD
ID
Material
DLW Internal Volume
Total delay volume
DLW option
0.50 mm
550 mm
FEP tubing
1/16 in. (1.58 mm)
108 μL
51 mm
22
Stainless steel
20.5 mm
0.72 mm (Gauge 22)
0.41 mm
Stainless steel
205 μL + Injection loop
volume (204.7)
Manifold: 90 μL
Holding Loop: 108 μL
Syringe Needle: 6.7 μL
Installed Injection Loop
Accela Open Autosampler Hardware Manual
1
Introduction
Open Autosampler Components
DLW-2 option
0.41 mm
855 mm
Stainless steel, passivated
0.72 mm (Gauge 22)
118 μL
Not applicable, integrated in
DLW-2 Holding Loop
Not applicable, integrated in
DLW-2 Holding Loop
208 μL + Injection loop
volume
Manifold: 90 μL
Holding Loop: 118 μL
Installed Injection Loop
5