Table 43. Tray type parameters
Item
Vial Trnsprt
ZSlideOffRetr
YSlideOffDist
Barcode
Z Tolerance
Z Retract
a
(Sheet 2 of 3)
Description
The mode to transport a vial is defined by:
• None = No transport required. Example liquid sampling.
• Magnet = Magnetic Transport. Requires magnetic vial caps.
The injection unit moves to the side to slide the vial off when
moved away.
• Needle = Device is transported with the syringe needle
inserted. Example: special vials where a magnetic cap cannot
be crimped, for example, Vacutainer.
Remark: This item became active beginning with PAL Firmware
level 2.5.x.
If magnetic transport is selected, the distance can be defined to
move the syringe slider up (Z-direction), lifting up the vial by
magnetic force. Measured in mm.
If magnetic transport is selected, the distance can be defined to
move the injection unit to the side, Y-direction, to slide the vial
off. Measured in mm.
If a Barcode Reader is installed, the mode has to be set for
AutoFix.
None = No Barcode Reader option requested for this tray type.
A tolerance window to give a plus/minus range (expressed in mm)
where the injection unit must expect an object.
If the value of this item is > 0, this is a relative detection mode and
the sensor from the injection unit needle guide is active.
If the value of this item is set to 0 the sensor of the needle guide is
turned off and the syringe slider moves to an absolute value, z-axis
position as specified in the object.
The distance to lift up the syringe slider before the y-, z-axes
assembly moves across the unit can be specified, expressed in mm.
A
Reference Information
Accela Open Autosampler Hardware Manual
Special Functions
129