Chapter 4 Operating The Temperature Controllable Stack Coolers; Temperature Setting, Control, And Alarm; Temperature Setting - Thermo Scientific Accela Open Autosampler Hardware Manual

Table of Contents

Advertisement

Operating the Temperature Controllable Stack
Coolers

Temperature Setting, Control, and Alarm

Temperature Setting

Thermo Scientific
This chapter describes how to set the temperature, temperature control, and alarm. It also
discusses temperature stability and condensation build-up.
Contents
Temperature Setting, Control, and Alarm
Temperature Stability
Condensation Build-Up
This section contains information about the temperature setting, temperature control, and
temperature alarm.
Turning the power on (|) at the power supply module puts the control unit into self-check
mode. The set value for the temperature appears after a few seconds.
The stack/tray cooler temperature is preset by the factory to 10 °C.
To change the value
1. Press P briefly (less than two seconds) until SP1 appears.
2. Use the increment/decrement keys to select the desired value.
3. Press P again.