Editing And Viewing Methods - Thermo Scientific Accela Open Autosampler Hardware Manual

Table of Contents

Advertisement

3
Operating the Accela Open Autosampler
Methods

Editing and Viewing Methods

58
Accela Open Autosampler Hardware Manual
To create a new method
1. Choose Menu > Methods > Select the Insertion Point > Instrument Method.
2. Enter a new method name.
After you create a copy of the method, the method parameters appear and can now be
edited. You can change the cycle and syringe entries.
3. If the method is new (that is, added), select and enter a cycle that is appropriate for the
application.
4. Select the specific syringe to be used by the method.
Note Once you create and save a method, you cannot change the cycle and syringe.
To use a different cycle or syringe, you must create a new method.
5. Assign parameter values according to the application requirements.
You can view and change method parameters (cycle and syringe not included) from the
Method menu as follows.
To edit or view a method
1. Scroll to and select the parameter you want to change. Assign the new value and press the
ENTER key.
2. Exit the parameter list by pressing either Home function key (F4) to return to the
top-level Job Queue menu or the ESC key to return to the previous menu.
3. To view method contents from the job queue displays, select the desired job, press
ENTER, and then press the View Method function key.
Thermo Scientific