Section F3-Setup/Objects - Thermo Scientific Accela Open Autosampler Hardware Manual

Table of Contents

Advertisement

A
Reference Information
Special Functions

Section F3-Setup/Objects

124
Accela Open Autosampler Hardware Manual
Use the Objects section opened in the Extended User mode to visualize new classes and
known classes that might contain new items accessible only at the Extended User level.
To open this level, use the following path
Figure 66. Selecting Objects in Extended User mode
Table 42. Objects in the Extended User mode
Object
Syringes
Vials
Tray Types
Tray Holders
Trays
Wash Stations
(Figure
66).
a
(Sheet 1 of 4)
Description
All loaded syringes are selectable. At the Utilities User Level only
the active syringe is visible.
The Firmware Objects accessed on the PAL Extended User mode
have more specific items selectable than at the User Level.
Tolerances or ranges such as Maximum Penetration can be defined
and limited for the user so that this value cannot be exceeded.
The object class Tray Type is only accessible via the PAL Extended
User Level. All loaded Tray Types are visible and selectable.
The Firmware Objects accessed on the PAL Extended User Level
have more specific items selectable than at the User Level.
Depending on the tray holder type, an Access and Restore Path or
a Heater and Agitator (motor) is assigned.
All loaded trays are visible and accessible. A tray must be assigned
to a tray holder. A TrayTypeGroup identification checks if the
specified tray type is allowed in this combination.
The path offset x, y, z is used only if a path is assigned to the
specified tray holder. Typically a stack or agitator uses a path.
The wash station and its positions, Wash1 and Wash2, are visible
and selectable.