Possible Positions; 3906 Light Module - Thermo Scientific 3906 Operating Manual

Environmental chamber
Table of Contents

Advertisement

Thermo Scientific

Possible Positions

One light module must be placed on top most position, otherwise the
temperature performance is influenced.

3906 Light Module

Light module 1 at the top position installed
shelf holder installed at first hole position from above
 distance from the chamber top = 2 inches ( 51 mm)
Standard position of shelf 1 (supreme shelf)
 shelf holder installed at 5
above
 distance from the chamber top = 10 inches (
250 mm)
distance between shelf 1 surface & lamps sur-
face = 6.9 inches (175 mm)
Further possible positions of shelf 1
(supreme shelf)
 shelf holder installed at 4
inches from the chamber top (but the
access port left side is at this height!)
Position of light module 2
 shelf holder installed at 7th hole position from
above
Standard position of shelf 2 (bottom shelf)
shelf holder installed at 11
distance from the chamber top = 22 inches ( 558 mm)
distance between shelf 2 surface & lamps surface = 6.9 inches (175 mm)
Further possible positions of shelf 2
shelf holder installed at 10
th
hole position from
th
hole position = 8
th
hole position from above
th
hole position = 20 inches from the chamber top
Product Description |
Chapter 5
Model 3900 Series | 5-13