Thermo Scientific Accela Open Autosampler Hardware Manual page 188

Table of Contents

Advertisement

B
PAL Loader
PAL Loader Software Operation
170
Accela Open Autosampler Hardware Manual
6. In PAL Loader software version 2.1.x, load or activate a file (with extension .sss) by
dragging it to the window
Figure 91. Drop a file into the Update window
7. Activate the process by clicking Start Update
The process is initialized and takes approximately 3–5 minutes.
Figure 92. Update process for serial communication completed
The RAM and flash memories are erased and replaced by the new data. When the process
is complete, this message appears: Update finished (from the task bar).
If there is a problem, a message similar to the following appears: Programming target
memory failed at 0x000A2D00.
The Update log entry memory failed ####.
(Figure
91).
(Figure
92).