Installing the Keypad Terminal
Thermo Scientific
Figure 13. Installing the keypad terminal
To install the keypad terminal
1. Install the safety shield on the left and right sides to the outside of the x axis. Use the
provided, longer thumbscrew on the side where you plan to install the keypad.
2. Install the keypad mounting bracket on either the right or the left side of the x axis
(Figure
13).
3. Connect one end of the white coiled cable to the keypad and the other end to the
TERMINAL (SER3) interface jack on the back side of the x axis
Figure 14. White coiled cable connection
4. Place the keypad terminal onto its mounting bracket.
2
Installing the Accela Open Autosampler
Installing the Autosampler and DLW
(Figure
14).