Thermo Scientific Accela Open Autosampler Hardware Manual page 150

Table of Contents

Advertisement

A
Reference Information
Special Functions
Trays
132
Accela Open Autosampler Hardware Manual
• Pattern Type, Staggrd–
The sample or well positions are arranged in a staggered pattern.
Offset of second Row is shifted by –50% of the hole pattern.
Figure 69. Pattern tray type of Staggered–
You can change the following object Tray Items by selecting the particular item
First select the desired tray, such as Tray1.
a
Table 44. Tray items
(Sheet 1 of 3)
Item
Tray Holder
TrayTypeGroup
Tray Type
Offset X
(Figure
69)
Description
Specifies the tray holder that the tray is assigned to.
Displays all tray type groups that are allowed for this particular
combination of tray and tray holder. The TrayTypeGroup for the
tray type itself is specified for each tray type.
Links the tray type to be used to the tray. This item is usually set at
the User Level in Utilities/Trays.
The X offset is defined as the distance from the center of the
teaching position from the tray holder to the edge of the tray in
X-direction. Measured in mm.
This item is available at the User Level in the section,
Utilities/Tray, and can be used for fine tuning a tray installed in a
tray holder. Important for well-plates 384 installed in a Stack.
Remark: Position#1 is the only correction point.
(Table
44).
Thermo Scientific