Reference Information
Special Functions
Wash Stations
134
Accela Open Autosampler Hardware Manual
a
Table 44. Tray items
(Sheet 3 of 3)
Item
dxCol
dyCol
dzCol
a
Items displayed in italics are available beginning with PAL Firmware level 4.1.x.
You can change the following object Wash Station items by selecting the particular item
(Table
45). First select the desired wash station, such as Wash1.
Table 45. Wash station items
Item
Position X
Position Y
Position Z
Z Tolerance
Z Retract
Max Penetrat
Needle Penetr
Description
Correction for inclination of a tray in X-column direction. For
details, see Section F, "Description and Installation," point 4.2.3.2
above.
Correction for inclination of a tray in Y-column direction. For
details, see Section F, "Description and Installation," point 4.2.3.2
above.
Correction for inclination of a tray in Z-column direction.
a
(Sheet 1 of 2)
Description
x-axis position for the wash station
y-axis position for the wash station
z-axis position for the wash station
A tolerance window for a plus/minus range (expressed in mm)
where the injection unit has to expect an object.
If the value of this item is > 0, a relative detection mode occurs
and the sensor from the injection unit needle guide is active.
If the value of this item is set to 0, the sensor of the needle guide is
turned off and the syringe slider moves to an absolute value, z-axis
position as specified in the object.
Specifies the distance to lift up the syringe slider before the y-,
z-axes assembly moves across the unit, expressed in mm.
Defines the maximum allowed Needle Penetration as a safety so
that the user cannot destroy the needle tip by going too far down.
The maximum needle penetration must not exceed the length of
the syringe needle, considering the loss due to the needle guide
and so on. Typical needle length is 51 mm; maximum penetration
must not exceed 48 mm.
This item is identical to that at the User Level. The value defined
at this level is mirrored in the Extended User Level.