Description of Object Positions

Tray Holders Stack or Stack Cooler with 3 or 6 Drawers
Thermo Scientific
Figure 33. Menu screen object Tray Holder
4. Select Position X and press ENTER.
The injection unit moves to the previously defined x-axis position.
5. Rotate the outer knob to adjust the x-axis position to the tray holder reference position.
6. Press the inner knob to enter the Position X value.
7. Repeat
step 4
through
If the F3 button (Move to Zero) is activated, the injection unit moves to the HOME
position.
8. Verify the defined x, y, z positions by pressing F1: Check Pos.
The release of PAL firmware level 4.1.2 makes possible the correction of all three axes, x, y,
and z, in the firmware class tray. You can correct the inclination for a tilted tray in the
direction of the row, the column, or both. Teaching is possible in the Utilities section by using
the path: Menu > Utilities > Trays.
This section contains a description of the object positions.
The reference position for a stack or stack cooler is identical. The reference position in the
case of a 3-drawer stack is a hole in the upper drawer and in a 6-drawer stack, in the second
drawer from the top (same height level), as shown in
6
for Position Y and Position Z.
2
Installing the Accela Open Autosampler
Accela Open Autosampler Object Positions
Figure
34.
Accela Open Autosampler Hardware Manual
41