Thermo Scientific Accela Open Autosampler Hardware Manual page 47

Table of Contents

Advertisement

Thermo Scientific
Preparing and Installing the DLW Syringe
This section contains information on how to prepare and install the DLW Syringe.
To prepare and install the DLW Syringe
1. Prepare the DLW Syringe by inserting the DLW Plunger Holder.
a. Move the plunger manually down to the stop position, and release a slight amount of
pressure to the plunger tip by pulling a fraction of a millimeter backwards.
b. Install the DLW Plunger Holder and tighten the Allen screw (Allen key #6) firmly
(Figure
23).
Figure 23. Inserting DLW Plunger Holder on DLW Syringe
Tip Make sure the syringes are primed before beginning sample preparation. Prime all
liquid syringes manually first before inserting them into the Open Autosampler
system.
2. Screw the prepared DLW Syringe into the holder. Hold the syringe at the lower metal
mount to tighten the syringe
2
(Figure
24).
Accela Open Autosampler Hardware Manual
Installing the Accela Open Autosampler
Installing the Autosampler and DLW
29