Wash Station - Thermo Scientific Accela Open Autosampler Hardware Manual

Table of Contents

Advertisement

Wash Station

Thermo Scientific
Table 13
lists the item you can select to change the injector configuration.
Table 13. Utility item
Item
Needle Penetr
Table 14
lists the functions available for Wash Station by pressing the appropriate function
key.
Table 14. Utility function keys for Wash Station
Function key
F3 Movto Wash
F4 HOME
Table 15
lists the items you can select to change the wash station configuration.
Table 15. Wash station items
Item
Needle Penetr
Rinse Time
Description
By selecting the Needle Penetr parameter, you can check or change
the injector needle penetration value. To ensure reproducible
sample injections and minimize carryover, make sure that the
needle penetration depth is accurately set.
Description
The injection unit moves to the selected wash station port. By
selecting the Needle Penetr parameter on the same screen, you can
check or change the wash station needle penetration value.
The injection unit moves to its HOME position and the Job
Queue menu is displayed.
Description
By selecting the Needle Penetr parameter, you can check or change
the wash station needle penetration value.
If rinse time is activated (value > 0), the solenoid of a fast or active
wash station opens for the specified time after the syringe needle
has been removed from the wash port (after completion of the
syringe wash cycle).
The solvent flows into the wash port without the restriction of the
needle; be aware of higher solvent consumption.
3
Operating the Accela Open Autosampler
Accela Open Autosampler Hardware Manual
Utility Functions
65