Changing the DLW Holding Loop
Thermo Scientific
With the DLW-2 option, a new DLW holding loop design has been introduced. The DLW-2
holding loop is a high performance loop, which is compatible with the previously used FEP
loop with a separate needle and flow diverter. The DLW-2 holding loop gives better
performance in regard to carryover and low volume injection.
Table 51
shows the differences between the DLW and DLW-2 holding loops.
Table 51. Comparison of the two holding loop designs (Sheet 1 of 2)
Part or Parameter
Holding loop design
Material:
OD:
ID:
Length:
Volume:
Syringe needle
Length:
Gauge:
Material:
DLW flow diverter
Length:
OD:
ID:
Material:
DLW holding loop
P/N 00950-01-00317
FEP tubing
1/16 in. (1.58 mm)
0.50 mm
550 mm
108 μL
P/N 00950-01-00328
51 mm
22
Stainless steel
P/N 00950-01-00319
20.5 mm
0.72 mm (Gauge 22)
0.41 mm
Stainless steel
DLW-2 holding loop
P/N 00950-01-00360
Loop stainless steel, passivated
0.72 mm (Gauge 22)
0.41 mm
855 mm
118 μL
Not applicable, integrated in
DLW-2 holding loop
Not applicable, integrated in
DLW-2 holding loop
Accela Open Autosampler Hardware Manual