Installing the DLW Option

Installing the DLW Wash Station
Thermo Scientific
The other end of the cable is connected to the wash station connector on the Control-xt
board, as shown in
Figure
Figure 19. Electrical connection from DLW pump holder to PAL-xt board
This section provides instructions on installing the various parts of the DLW option.
Installing the DLW Wash Station
Installing the DLW Syringe Holder Assembly
You can connect the DLW Wash Station anywhere along the x axis. However, Thermo Fisher
Scientific strongly recommends that you install it as close as possible to the injection valve.
Make the travel path of the injection unit from the valve to the wash station as short as
possible to keep the cycle time as brief as possible.
Figure 20
shows two configurations for installing DLW modules, favoring a short connection
from the injection valve to the other HPLC system components on the one hand, and always
having the DLW Wash Station close to the injection valve on the other. Always attach the
DLW Pump Module, with the connected DLW tubing from the kit, on the far left side of the
x axis.
Such a configuration also has the advantage that you can connect the waste tube from the
injection valve to the front waste inlet of the wash station module. This waste tube is part of
the DLW option kit.
19.
Accela Open Autosampler Hardware Manual
2
Installing the Accela Open Autosampler
Installing the Autosampler and DLW
25