Thermo Scientific Accela Open Autosampler Hardware Manual page 202

Table of Contents

Advertisement

C
PAL Object Manager
Getting Started with PAL Object Manager
Starting the PAL Object Manager Application
Using the Object Manager Application
184
Accela Open Autosampler Hardware Manual
To start the PAL Object Manager application
1. From the desktop, click the PAL Object Manager icon (
The application opens and displays the various user buttons. The main part of the
application shows the Object List folder pane with the current Object Lists folder.
All objects must correspond with the proper version of the firmware as identified by
Thermo Scientific. For example, for objects using 4.x.x firmware, valid revisions are A, B,
or C. For objects using a previous version of the software, use revision K.
The status bar displays the previously used communication mode.
2. To see the version information, choose Help > About.
The software displays the Object Manager software version.
The main purpose of the PAL Object Manager software is to load PAL firmware object lists
on a system that has added a hardware module. To add an object list, follow this procedure,
which describes adding a syringe object to an existing Accela Open Autosampler.
).