Thermo Scientific Accela Open Autosampler Hardware Manual page 146

Table of Contents

Advertisement

A
Reference Information
Special Functions
Tray Type
128
Accela Open Autosampler Hardware Manual
You can change Tray Type items by selecting the particular item
desired Tray Type, such as VT32-20.
Table 43. Tray type parameters
Item
Row Length X
Row Length Y
Col Length X
Col Length Y
Spl Per Row
Spl Per Col
Pattern
TrayTypeGroup
Plate Thickn
Vial Height
a
(Sheet 1 of 3)
Description
Defines the Row length in the X-direction from the center of the
first to the center of the last vial position, measured in mm.
Defines the Row length in the Y-direction from the center of the
first to the center of the last vial position, measured in mm.
Defines the Column length in the X-direction from the center of
the first to the center of the last vial position, measured in mm.
Defines the Column length in the Y-direction from the center of
the first to the center of the last vial position, measured in mm.
Number of samples per Row.
Number of samples per Column.
The pattern of the Row or Column arrangement has to be
defined.
• Regular = square arrangement (standard)
• Staggrd+ = Rows are arranged in a staggered pattern.
Offset of second Row is shifted by +50% of hole pattern.
• Staggrd- = Rows are arranged in staggered pattern.
Offset of second Row is shifted by –50% of hole pattern.
Software protection so that not every tray type can be placed on
any tray holder. Example: You cannot fit VT32-20 into a Stack.
Thickness of tray plate bottom (not the total height of the tray).
Measured in mm.
Total height of the vial including cap to seal if applicable.
Measured in mm.
Remark: The sum of the plate thickness and vial height adds up to
the total height This is the point where the injection unit expects
an object.