If there is an FEP holding loop installed, remember to remove now the flow diverter from
the DLW Manifold
(Figure
DLW-2 holding loop is installed because the flow diverter is part of the holding loop.
Figure 116. Remove the flow diverter
8. Insert the DLW-2 holding loop to the DLW Manifold and tight the ferrule.
9. Fix the DLW-2 holding loop with the slotted needle adapter to the DLW-2 Syringe
Holder Assembly. Make sure the opening of the needle adapter is facing back.
10. Fix the DLW-2 holding loop to the needle adapter with the retaining nut.
11. Insert the DLW-2 syringe adapter without installed syringe
Figure 117. Install the DLW-2 syringe adapter
116). Removing the flow diverter is not necessary if the
Accela Open Autosampler Hardware Manual
E
Changing the DLW Holding Loop
(Figure
117).