Thermo Scientific Accela Open Autosampler Hardware Manual page 223

Table of Contents

Advertisement

Thermo Scientific
If there is an FEP holding loop installed, remember to remove now the flow diverter from
the DLW Manifold
(Figure
DLW-2 holding loop is installed because the flow diverter is part of the holding loop.
Figure 116. Remove the flow diverter
8. Insert the DLW-2 holding loop to the DLW Manifold and tight the ferrule.
9. Fix the DLW-2 holding loop with the slotted needle adapter to the DLW-2 Syringe
Holder Assembly. Make sure the opening of the needle adapter is facing back.
10. Fix the DLW-2 holding loop to the needle adapter with the retaining nut.
11. Insert the DLW-2 syringe adapter without installed syringe
Figure 117. Install the DLW-2 syringe adapter
116). Removing the flow diverter is not necessary if the
Accela Open Autosampler Hardware Manual
E
Changing the DLW Holding Loop
(Figure
117).
205