Set TV Line Number for Synchronization
:TRIGger[:SEQuence]:TV:LINE <line>
:TRIGger[:SEQuence]:TV:LINE?
This command allows you to set the TV line number to which the
analyzer will synchronize its sweep.
Factory Preset
and *RST:
17
Range:
The range is dependent upon the Field Mode, as
described in the command
:TRIGger[:SEQuence]:TV:FMODe ENTire|ODD|EVEN .
The minimum value is the minimum line, and rolls over
to the maximum value. The maximum value is the
maximum line, and rolls over to the minimum value.
Remarks:
Refer to the command
:TRIGger[:SEQuence]:SOURce:TV, which is used to
activate TV triggering.
Set Analyzer for TV Picture Monitoring
:TRIGger[:SEQuence]:TV:MONitor OFF|ON|0|1
:TRIGger[:SEQuence]:TV:MONitor?
The currently-selected standard is used to determine the proper setup
of the hardware state of the video digitization circuitry for presentation
of the TV picture.
Factory Preset
and *RST:
Off
Front Panel
Access:
Trig, TV Trig Setup,TV Monitor
Set the Video Waveform Sync. Pulse Direction
:TRIGger[:SEQuence]:TV:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:TV:SLOPe?
This command defines the direction of the sync pulse on the TV video
waveform.
Factory Preset
and *RST:
POSitive
Chapter 5
Language Reference
TRIGger Subsystem
5-147