Mixer Bias will be Off if AUTO Harmonic and Presel Mixer Type is
NOTE
selected.
Front Panel
Access:
Set External Mixer LO Harmonic Value
[:SENSe]:MIXer:HARMonic <value>
[:SENSe]:MIXer:HARMonic?
This command allows you to set the LO harmonic value for mixers
other than the HP/Agilent 11970-Series or 11974-Series Mixers. This is
done after the mixer harmonic configuration
([:SENSe]:MIXer:HARMonic:AUTO <boolean>) is set to manual.
Factory Preset
and *RST:
Range:
Remarks:
Set External Mixer LO Harmonic Mode
[:SENSe]:MIXer:HARMonic:AUTO <boolean>
[:SENSe]:MIXer:HARMonic:AUTO?
This command allows you to set the external mixer LO harmonic mode
to either automatic or manual. The manual mode is used with mixers
other than the HP/Agilent 11970-Series or 11974-Series Mixers to
manually choose the required LO harmonic.
Factory Preset
and *RST:
Chapter 5
The bias signal is present on the center conductor of the
IF Input connector on the front panel. See related
command [:SENSe]:MIXer:BIAS <numeric>.
Input/Output (or Input), Input Mixer, Mixer Config, Mixer
Bias On Off
–8 (Band A, 26.5-40 GHz)
Any non-zero integer from –50 to 50, inclusive
The harmonic value with its associated sign is
automatically determined from the external mixer
band selected. This is the AUTO mode. For mixers
other than the HP/Agilent 11970-Series or
11974-Series, an LO harmonic other than that
determined in the AUTO mode may be required. This is
achieved using this command and related command
[:SENSe]:MIXer:HARMonic:AUTO <boolean>.
AUTO