<ampl> –120 dBm to +100 dBm
<connected> 0 or 1
Front Panel
Access:
Display, Limits, Properties, X Axis Units Freq Time
Delete Limit Line
:CALCulate:LLINe[1]|2:DELete
Deletes the selected limit line.
Display the Limit Line
:CALCulate:LLINe[1]|2:DISPlay OFF|ON|0|1
:CALCulate:LLINe[1]|2:DISPlay?
Controls the display of the current limit line.
Factory Preset
and *RST:
Off
Front Panel
Access:
Display, Limits, Modify, Limit On Off
Test the Data Against the Limit Line
:CALCulate:LLINe[1]|2:FAIL?
Queries the status of the limit line testing. Returns a 0 if the data
passes, and returns a 1 if there is a failure. This query only makes
sense if margin or limit test is On. Use the command
:CALCulate:LLINe[1]|2:STATe OFF|ON|0|1 to activate limit line
testing.
Set the Margin Size
:CALCulate:LLINe[1]|2:MARGin <ampl_rel>
:CALCulate:LLINe[1]|2:MARGin?
Allows you to define the amount of measurement margin that is added
to the designated limit line.
Factory Preset
and *RST:
not affected
Default Units: dB
Remarks:
The margin must be negative for upper limit lines, and
positive for lower limits.
Front Panel
Access:
Display, Limits, Modify, Margin On Off
Chapter 5
Language Reference
CALCulate Subsystem
5-17