Language Reference
SENSe Subsystem
This command allows the selection of one of the pre-defined bands
corresponding to the external mixer currently in use.
Factory Preset
and *RST:
Remarks:
Bands K, E, W, F, D, G, Y, and J are not available if Mixer Type is set to
NOTE
Presel.
Front Panel
Access:
External Mixer Bias Adjust
[:SENSe]:MIXer:BIAS <numeric>
[:SENSe]:MIXer:BIAS?
This command allows the adjustment of an internal bias source for use
with external mixers.
Factory Preset
and *RST:
Range:
Default Unit:
Remarks:
Mixer Bias will be Off if AUTO Harmonic and Presel Mixer Type is
NOTE
selected.
Front Panel
Access:
Set External Mixer Bias On/Off
[:SENSe]:MIXer:BIAS:STATe <boolean>
[:SENSe]:MIXer:BIAS:STATe?
This command activates an internal bias source for use with external
mixers.
Factory Preset
and *RST:
5-98
Band A (26.5-40 Ghz)
If the mixer harmonic configuration
([:SENSe]:MIXer:HARMonic:AUTO <boolean>) is
set to manual, then a query will return "USER."
Input/Output (or Input), Input Mixer, Ext Mix Band
0
–10 mA to 10 mA
mA
The bias signal is present on the center conductor of the
IF INPUT connector on the front panel. See related
command [:SENSe]:MIXer:BIAS[:STATe]
OFF|ON|0|1.
Input/Output (or Input), Input Mixer, Mixer Config, Mixer
Bias On Off
Off
Chapter 5