Trace Y-Axis Reference Level; Trace Y-Axis Reference Level Offset - Agilent Technologies ESA-E Series Programmer's Manual

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
DISPlay Subsystem
Range:
1 kHz to 240 kHz
Default Unit:
Hz
Front Panel
Access:
AMPLITUDE/Y Scale, Scale/Div

Trace Y-Axis Reference Level

:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel <ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel?
Sets the amplitude value of the reference level for the y-axis.
Factory Preset
and *RST:
0 dBm
Range:
With zero reference level offset:
ESA E4401B, E4411B: –327.6 to 50 dBm
ESA E4402B, E4403B: –327.6 to 55 dBm
ESA E4404B: –327.6 to 55 dBm
ESA E4405B: –327.6 to 55 dBm
ESA E4407B, E4408B: –327.6 to 55 dBm
–149.9 to 55 dBm with zero reference level offset and
max mixer level = –10 dBm. In external mixing, the
range is –327.5 to –10 dBm.
Default Unit:
current active units
Remarks:
The input attenuator setting may be affected. The
minimum displayed value of reference level is
–327.6 dBm, and the maximum displayed value is
327.6 dBm. See the remarks given for the command
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet
<rel_ampl>
Front Panel
Access:
Amplitude Y Scale, Ref Level

Trace Y-Axis Reference Level Offset

:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet <rel_ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet?
Sets the amplitude level offset for the y-axis.
Factory Preset
and *RST:
0 dB
Range:
–327.6 to 327.6 dB
5-44
Chapter 5

Hide quick links:

Advertisement

Table of Contents
loading