Preparation for Use

In This Chapter

Before turning the 4339B ON, you must rst set the 4339B to match the available power LINE
voltage.
If the 4339B's power LINE voltage and frequency are properly set and ready to use, you can
skip this chapter.

Power Requirements

The 4339B's power source requirements are as follows:
LINE Voltage : 100 / 120 / 220 / 240 V ac (
LINE Frequency : 47 to 66 Hz
Power Consumption : 45 VA maximum

To Set Power LINE Voltage

1. Con rm that the power cable is disconnected.
2. Slide the LINE Voltage selector on the rear panel to match the power LINE voltage which
will be used (see Table 1-1).
Voltage Selector
6
Table 1-1. Power Voltage Selector Setting
Line Voltage
6
100V/120Vac(
10%)
6
220V/240Vac(
10%)
10%)
Required Fuse
UL/CSA type, Time delay 0.5A 250V
(Agilent part number 2110-0202)
UL/CSA type, Time delay 0.25A 250V
(Agilent part number 2110-0201)
1
Preparation for Use
1-1