Resolution Bandwidth Automatic
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO OFF|ON|0|1
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO?
Couples the resolution bandwidth to the frequency span.
Factory Preset
and *RST:
On
Video Bandwidth
[:SENSe]:BANDwidth|BWIDth:VIDeo <freq>
[:SENSe]:BANDwidth|BWIDth:VIDeo?
Specifies the video bandwidth.
Factory Preset
and *RST:
3 MHz
Range:
1 Hz to 3 MHz. This range is dependent upon the
setting of
[:SENSe]:BANDwidth|BWIDth[:RESolution].
Default Unit:
Hz
Front Panel
Access:
BW/Avg, Video BW Auto Man
Video Bandwidth Automatic
[:SENSe]:BANDwidth|BWIDth:VIDeo:AUTO OFF|ON|0|1
[:SENSe]:BANDwidth|BWIDth:VIDeo:AUTO?
Couples the video bandwidth to the resolution bandwidth.
Factory Preset
and *RST:
On
Front Panel
Access:
BW/Avg, Video BW Auto Man
Video to Resolution Bandwidth Ratio
[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio <number>
[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio?
Specifies the ratio of the video bandwidth to the resolution bandwidth.
Factory Preset
and *RST:
1.0
Range:
0.00001 to 3.0e6
Front Panel
Access:
BW/Avg, VBW/RBW Ratio