Error Messages
Greater than 1000: Personality Specific Error Messages
10161
Lower Custom Mask is Invalid!
The user-specified lower custom mask cannot be
resolved into a limit line.
10231
Measurement failed for unknown reasons.
Check instrument settings and restart measurement.
10154
Measurement not defined for Out of Band.
User has changed to an out-of-band frequency range.
The band measurement only operates in the selected
band.
10227
Measurement suspended until carrier is turned
off.
The receive channel power and the receive spur
measurements are specified with the attenuation set to
0 dB. To prevent overload, the frequency spectrum of
interest is monitored for signal levels which exceed a
specified threshold before setting the attenuator to
0 dB. If a carrier is found, this message is displayed and
the completion of the measurement will not occur until
the carrier is removed. The carrier check may be turned
off using the properties form under the front-panel
Mode Setup
threshold which determines a carrier on the properties
form.
10155
No Fast ADC hardware installed. Meas
unavailable.
The analyzer cannot use sweeptimes less than 5 msec
when (Option B7D or Option AYX) is not installed.
Therefore the measurement will not be executed.
10147
Opt B7D bootrom requires upgrade.
The (Option B7D) bootrom revision is not supported by
the curently loaded personality version. Refer to the
user's guide for the personality in use.
7-40
key. You may also change the signal
Chapter 7