Trigger Offset On/Off; Trigger Source - Agilent Technologies ESA-E Series Programmer's Manual

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Front Panel
Access:
Trig, Trig Offset

Trigger Offset On/Off

TRIGger:SEQuence:OFFSet:STATe ON|OFF
TRIGger:SEQuence:OFFSet:STATe?
This command enables or disables trigger offset.
Factory Preset
and *RST:
Off
Example:
:TRIG:SEQ:OFFS
Range:
On/Off
Remarks:
Trigger offset can only be turned on when in zero span
and the resolution bandwidth is 1 kHz or greater
(non-digital bandwidths). Trigger offset is available for
all trigger modes.
Front Panel
Access:
Trig, Trig Offset

Trigger Source

:TRIGger[:SEQuence]:SOURce IMMediate|VIDeo|LINE|
EXTernal|TV
:TRIGger[:SEQuence]:SOURce?
Specifies the source (or type) of triggering used to start a measurement.
Immediate is free-run triggering
Video triggers on the video signal level
Line triggers on the power line signal
External allows you to connect an external trigger source
TV triggers on the selected line of a TV frame
Factory Preset
and *RST:
Immediate (free-run triggering)
Front Panel
Access:
Trig, Free Run
Trig, Video
Trig, Line
Trig, External Pos Neg
Trig, TV
Chapter 5
Language Reference
TRIGger Subsystem
5-145