729
In <filename>: error reading file
Appears when loading data from a limit line or
corrections disk file and a failure to the file occurs.
730
In <filename>, line <numeric_value>: line too
long
When loading data from a limit line or corrections disk
file, this message will appear if the length of any line in
the file exceeds 255 characters.
731
In <command>: bad data count
(<numeric_value>): expected multiple of
<numeric_value>
This message indicates that the data sent to a
corrections or limit table via the
commands does not have the expected length for the
table. For example, this message would appear if an
attempt were made to merge 7 numeric values into a
limit table, since each logical entry requires 3 values
(frequency, amplitude, and connected).
732
In <filename>, line <numeric_value>: error
parsing tokens
This message may appear when loading data from a
limit line or corrections disk file. It indicates a problem
in the attempt to break a string of text into tokens.
There may be too few tokens in the string. This
typically happens when there are too few numeric
values in the [
733
In <filename>, line <numeric_value>: <xxx> is
not numeric
This message may appear when loading data from a
limit line or corrections disk file. It indicates that a
non-numeric token <xxx> was found where a numeric
token was expected.
735
In <filename>: bad amplitude unit <unit>
This message indicates that unit <unit> is not
recognized or supported.
Chapter 7
201 to 799: Device-Specific Errors
DATA
] section of a limit or corrections file.
DATA
Error Messages
or
MERGE
7-29