Agilent Technologies ESA-E Series Programmer's Manual page 244

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Factory Preset
and *RST:
100 kHz
Range:
5 kHz to 1.2 MHz
Default Unit:
Hz
Front Panel
Access:
Det/Demod, Demod, FM Deviation
Demodulation Control
[:SENSe]:DEMod:STATe OFF|ON|0|1
[:SENSe]:DEMod:STATe?
Turns demodulation on or off.
Factory Preset
and *RST:
Off
Front Panel
Access:
Det/Demod, Demod, Off
Demod Time
[:SENSe]:DEMod:TIME <time>
[:SENSe]:DEMod:TIME?
Sets the time used for frequency domain demodulation.
Factory Preset
and *RST:
500 ms
Range:
2 ms to 100 s
Default Unit:
seconds
Front Panel
Access:
Det/Demod, Demod, Demod Time
Demod View
[:SENSe]:DEMod:VIEW[:STATe] OFF|ON|0|1
[:SENSe]:DEMod:VIEW[:STATe]?
This command causes the demodulated signal to be displayed. If FM
Demod is on, then the display scales the y-axis in units of kHz. The
scale/div is set with the command :DISPlay:WINDow:TRACe:Y
[:SCALe]:PDIVision:FREQuency <freq> if FM Demod is on. If FM
Demod is on, then several functions are not available; these include:
Log/Lin (display is always in linear), Y-Axis Units, Marker Search
functions, Normalize, Display Line, Peak Excursion, and Peak
Threshold. There is no effect when AM demodulation is used (only
applicable for FM demodulation).
Chapter 5
Language Reference
SENSe Subsystem
5-89

Hide quick links: