Language Reference

COUPle Subsystem

COUPle Subsystem
Some measurement settings are automatically coupled together to
optimize speed and accuracy. These commands control that coupling.

COUPle the Function to Other Settings

:COUPle ALL|NONE

:COUPle?

The instrument can automatically couple instrument settings together
for accurate measurements and optimum dynamic range. This
command is used to override the coupling for special measurement
needs.
Factory Preset
and *RST:
All
Remarks:
The following list of analyzer functions can either be
automatically coupled, or not coupled (manually set):
Residual bandwidth
Video bandwidth
Sweep time
Center frequency step
Attenuation
Marker count resolution
Source attenuation
Source amplitude step
COUPle NONE puts these functions into the manually
set (not coupled) mode. COUPle ALL puts the functions
into the auto coupled mode, and also puts the sweep
coupling mode into SA (couple all).
Front Panel
Access:
Auto Couple
5-38