Power On Type
:SYSTem:PON:TYPE PRESet|LAST
:SYSTem:PON:TYPE?
Sets the defined instrument conditions after a power-on or
PRESet - The instrument settings at power-on will be either the factory
preset or user preset, as set by :SYSTem:PRESet:TYPE FACTory|USER.
LAST - The instrument settings at power-on will be the settings at the
time of power down.
Factory Preset
and *RST:
The factory default is Preset. This parameter is
persistent, which means that it retains the setting
previously selected, even through a power cycle.
Front Panel
Access:
System, Power On/Preset, Power On Last Preset
Enable IF/Video/Sweep Output Ports
:SYSTem:PORTs:IFVSweep:ENABle OFF|ON|0|1
:SYSTem:PORTs:IFVSweep:ENABle?
This command enables or disables the IF, video, and sweep output ports
for Agilent ESA analyzers having options A4J (IF, Sweep, and Video
Ports) and AYX (Fast Time Domain Sweeps).
Factory Preset
and *RST:
On
Example:
:SYST:PORT:IFVS:ENAB ON
Range:
On/Off
Remarks:
Disable the output ports for faster measurement times.
Preset
:SYSTem:PRESet
Returns the instrument to a set of defined conditions. The particular set
is selected by :SYSTem:PRESet:TYPE. This command does not change
any persistent parameters. The term persistent means that the
command retains the setting previously selected, even through a power
cycle.
Front Panel
Access:
Preset
Chapter 5
Language Reference
SYSTem Subsystem
.
Preset
5-133