Agilent Technologies ESA-E Series Programmer's Manual page 251

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
SENSe Subsystem
Front Panel
Access:
Meas Setup, Avg Number On
Turn On or Off Harmonic Measurement Averaging
[:SENSe]:HARMonics:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:HARMonics:AVERage[:STATe]?
Turn on or off harmonic measurement averaging.
Factory Preset
and *RST:
Off
Front Panel
Access:
Meas Setup, Avg Number
Set Number of Measured Harmonics
[:SENSe]:HARMonics:NUMBer <integer>
[:SENSe]:HARMonics:NUMBer?
Set the number of harmonics to be measured.
Factory Preset
and *RST:
10
Range:
2 to 10, limited by the maximum frequency range of the
analyzer
Remarks:
This command sets the number of harmonics to
measure before computing the total harmonic
distortion. The minimum number is two (the
fundamental and the second harmonic).
Front Panel
Access:
Meas Setup, Harmonics
Set Harmonic Measurement Sweep Mode
[:SENSe]:HARMonics:SWEeptime:AUTO OFF|ON|0|1
[:SENSe]:HARMonics:SWEeptime:AUTO?
Set auto sweep time per harmonic to off or on.
Factory Preset
and *RST:
On
Remarks:
See [:SENSe]:HARMonics:SWEeptime <time>
Front Panel
Access:
Meas Setup, ST/Harmonic
Set Harmonic Measurement Sweep Time
[:SENSe]:HARMonics:SWEeptime <time>
5-96
Chapter 5