This command determines which bits in the Questionable Power
Condition Register will set bits in the Questionable Power Event
register, which also sets the Power Summary bit (bit 3) in the
Questionable Register. The variable <number> is the sum of the
decimal values of the bits you want to enable.
Factory Preset
and *RST:
Range:
Questionable Power Event Query
:STATus:QUEStionable:POWer[:EVENt]?
This query returns the decimal value of the sum of the bits in the
Questionable Power Event register.
The register requires that the equivalent PTR or NTR filters be set
NOTE
before a condition register bit can set a bit in the event register.
The data in this register is latched until it is queried. Once queried, the
data is cleared.
Questionable Power Negative Transition
:STATus:QUEStionable:POWer:NTRansition <number>
:STATus:QUEStionable:POWer:NTRansition?
This command determines which bits in the Questionable Power
Condition register will set the corresponding bit in the Questionable
Power Event register when that bit has a negative transition (1 to 0).
The variable <number> is the sum of the decimal values of the bits that
you want to enable.
Factory Preset
and *RST:
Range:
Questionable Power Positive Transition
:STATus:QUEStionable:POWer:PTRansition <number>
:STATus:QUEStionable:POWer:PTRansition?
This command determines which bits in the Questionable Power
Condition register will set the corresponding bit in the Questionable
Power Event register when that bit has a positive transition (0 to 1).
The variable <number> is the sum of the decimal values of the bits that
you want to enable.
Chapter 5
32767 (all 1's)
0 to 32767
0
0 to 32767
Language Reference
STATus Subsystem