Sweep Subsystem; Sweep Points - Agilent Technologies ESA-E Series Programmer's Manual

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

SWEEp Subsystem

This subsystem is used to set the controls and parameters associated
with the overall system sweep functions.

Sweep Points

:SWEEp:POINts <number of points>
:SWEEp:POINts?
This command sets the number of sweep points.
Factory Preset
and *RST:
401
Example:
:SWE:POIN 401
Range:
101 to 8192
Remarks:
This command is only available on analyzers with
firmware revision A.04.00 and later. The number of
sweep points can be set only with Agilent ESA models
E4401B, E4402B, E4404B, E4405B, and E4407B. The
number of sweep points may be queried on all
anlayzers with firmware revision A.04.00 and later.
Whenever the number of sweep points change, the
following functions are affected:
• All trace data is erased
• Any traces in view mode will go to blank mode
• Sweep time is re-calculated
• Any limit lines that are on will be turned off
Front Panel
Access:
Sweep, Points
Chapter 5
Language Reference
SWEEp Subsystem
5-125