Default Unit:
Hz
Front Panel
Access:
FREQUENCY/Channel, Start Freq
Stop Frequency
[:SENSe]:FREQuency:STOP <freq>
[:SENSe]:FREQuency:STOP?
Set the stop frequency.
Factory Preset
and *RST:
ESA E4401B, E4411B: 1.5 GHz
ESA E4402B, E4403B: 3.0 GHz
AESA E4404B: 6.7 GHz
ESA E4405B: 13.2 GHz
ESA E4407B, E4408B: 26.5 GHz
Range:
ESA E4401B, E4411B: –80 MHz to 1.58 GHz
ESA E4402B, E4403B: –80 MHz to 3.10 GHz
ESA E4404B: –80 MHz to 6.78 GHz
ESA E4405B: –80 MHz to 13.3 GHz
ESA E4407B, E4408B: –80 MHz to 27.0 GHz
Default Unit:
Hz
Front Panel
Access:
FREQUENCY/Channel, Stop Freq
HARMonics Subsection
[:SENSe]:
Harmonic Measurement Averages
[:SENSe]:HARMonics:AVERage:COUNt <integer>
[:SENSe]:HARMonics:AVERage:COUNt?
Set the number of averages for the harmonic measurement.
Factory Preset
and *RST:
10
Range:
1 to 1,000
Remarks:
This command specifies the number of sweep averages
over which the amplitude of each harmonic and the
total harmonic distortion will be calculated.
Intermediate averaged results will be displayed.
Chapter 5
Language Reference
SENSe Subsystem
5-95