Set Limit Line - Agilent Technologies 4396B Gpib Programming Manual

Network/spectrum/impedance analyzer
Hide thumbs Also See for 4396B:
Table of Contents

Advertisement

Lines 60 and 70 print the limit table heads on the BASIC SCREEN.
Line 90 de nes the table (array
the sweep parameter, the upper limit value, and the lower limit value as follows:
Segment
1
2
. .
.
Lines 110 to 130 call the subroutine
you de ned in line 100 (the analyzer can retain up to 18 segments).
d
The loop, lines 150 to 210, determines if you want to edit the table and con rms that the
segment is in the table.

Set Limit Line

In this portion of the program, the limit table (edited using BASIC) is transferred to the
analyzer.
Sweep Parameter
. .
.
) used to hold the limit values. It contains
Upper Limit
. .
.
(line 610) to edit and print as many segments as
Start to set the limit line
Clear existing limit table
Start to set a segment
Enter sweep parameter
Enter upper limit
Enter lower limit
Complete the segment
Complete the limit line
Set limit line ON
Programming Limit T est from Remote
Lower Limit
. .
.
a
8-3