10251
Error reading file: RHOPMCO. Please reinstall
cdmaOne.
The file is missing or corrupt. Please reinstall the
cdmaOne personality.
10245
Error reading file: SPCLIMIT.CSV. Cannot use
custom limits.
The file could be missing or corrupt. Create a new
limits file. Alternatively, the actual limits defined in the
file might not allow the measurement to be executed.
Redefine the limits or use the default limits. Restart
the measurement.
10180
Gate option not installed. Results may not be
accurate.
This measurement method requires the use of the
time-gate (option 1D6) in order to gate the spectrum
during the 50-90% part of the burst. If the gate option
is not installed, the measurement will still run
although this warning will be displayed.
10218
Hardkeys are disabled.
Some of the forms (for example Receiver Spurious in
GSM) do not allow the user to close the form without
either formally accepting or cancelling the form
settings. For this reason, all of the hardkeys are
disabled until the user terminates the form.
10233
Level is low, results may degrade.
The signal being measured is of low power. The results
may not be as accurate as they would be if the signal
was stronger.
10152
Lost trigger, aborting measurement.
The selected trigger source was present at the start of
the measurement, but timed out before the
measurement completed.