Error Messages
201 to 799: Device-Specific Errors
224
Option not licensed.
The selected option requires a license. Refer to the
installation procedures in the user's guide available for
this particular option.
209
Preselector centering failed
An attempt to center the preselector failed.
704
Printer interface error
An error occurred while trying to print. Make sure the
printer is turned on and properly connected.
705
Printer Type is None
The current printer type is set to
operations are possible. Change the type in the
Print Setup
211
RBW limited to 1kHz when Span > 5MHz
In spans greater than 5 MHz, narrow (digital)
resolution bandwidths, below 1 kHz, are not available.
217
RS-232 Interface Error
An error occurred on the serial interface.
213
Span limited to 5MHz when RBW < 1kHz
In narrow (digital) resolution bandwidths, below 1 kHz,
spans greater than 5 MHz are not available.
771
Store Ref trace before turning on Normalize
A reference trace must be available for the Normalize
function to be activated. Refer to the Agilent
Technologies ESA Spectrum Analyzers User's Guide,
where the
7-32
menu and try again.
key function is explained in detail.
Normalize
, so no print
None