Language Reference
SENSe Subsystem
Factory Preset
and *RST:
Off
Remarks:
When a measurement under the front panel
key is started, this command is turned off for video
averaging ([:SENSe]:AVERage:TYPE LPOWer). If this
command is turned on for video averaging when any of
the
MEASURE
measurement will be stopped.
Front Panel
Access:
BW/Avg, Average On Off
Type of Averaging for Measurements
[:SENSe]:AVERage:TYPE LPOWer|POWer
[:SENSe]:AVERage:TYPE?
This command selects the type of averaging to be performed.
LPOWer logarithmically averages the power of the video data (typical
units are dBm). This command is equivalent to pressing front panel
keys
BW/Avg, Average Type, Video.
POWer averages the linear power of successive measurements
(typical units are watts).
Factory Preset
and *RST:
LPOWer
Front Panel
Access:
BW/Avg, Average Type, Video Power

[:SENSe]:BANDwidth Subsection

Resolution Bandwidth
[:SENSe]:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:BANDwidth|BWIDth[:RESolution]?
Specifies the resolution bandwidth.
Factory Preset
and *RST:
3 MHz
Range:
10 Hz to 5 MHz with Option 1DR, narrow resolution
bandwidth; 1 kHz to 5 MHz without Option 1DR.
Default Unit:
Hz
Front Panel
Access:
BW/Avg, Resolution BW Auto Man