Agilent Technologies ESA-E Series Programmer's Manual page 279

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
STATus Subsystem
Factory Preset
and *RST:
32767 (all 1's)
Range:
0 to 32767
Questionable Positive Transition
:STATus:QUEStionable:PTRansition <number>
:STATus:QUEStionable:PTRansition?
This command determines which bits in the Questionable Condition
register will set the corresponding bit in the Questionable Event
register when that bit has a positive transition (0 to 1). The variable
<number> is the sum of the decimal values of the bits that you want to
enable.
Factory Preset
and *RST:
32767 (all 1's)
Range:
0 to 32767
5-124