Error Queues - Agilent Technologies ESA-E Series Programmer's Manual

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Error Queues

When a user-error condition occurs in the instrument as a result of
SCPI (remote interface) activity, it is reported to both the front-panel
display error-queue and the SCPI error queue. If it is a result of
front-panel activity, it reports to the front panel display error queue,
and may also report to the SCPI error queue depending on the error.
These two queues are viewed and managed separately.
Table 7-1
Characteristics of the Error Queues
Characteristic
Capacity (number of errors)
Overflow Handling
Viewing Entries
Clearing the Queue
Chapter 7
Front-Panel Display
Error Queue
11
Circular (rotating).
Drops oldest error as new
error comes in.
Press: System, Show Errors
Press: System, Show Errors,
Clear Error Queue
Error Messages
SCPI Remote Interface
Error Queue
30
Linear, first-in/first-out.
Replaces newest error with:
–350,Queue overflow
Use SCPI query
SYSTem:ERRor?
Power up.
Send a
command.
*CLS
Read last item in the queue.
Error Queues
7-13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents