Language Reference
SYSTem Subsystem
Persistent State Reset
:SYSTem:PRESet:PERSistent
Sets the persistent state values to their factory defaults. The term
persistent means that the command retains the setting previously
selected, even through a power cycle. Examples of persistent functions
are: GPIB address, power-on type, and preset type.
Front Panel
Access:
System, Restore Sys Defaults
Save User Preset
:SYSTem:PRESet[:USER]:SAVE
Saves the current instrument conditions as the user preset condition.
Front Panel
Access:
System, Power On/Preset, Save User Preset
Preset Type
:SYSTem:PRESet:TYPE FACTory|USER
Selects the preset state to be either factory-defined or user-defined
preset conditions.
Factory Preset
and *RST:
The factory default is FACTory. This parameter is
persistent, which means that it retains the setting
previously selected, even through a power cycle.
Remarks:
:SYSTem:PRESet:USER:SAVE defines the user preset.
Front Panel
Access:
System, Power On/Preset, Preset Factory User
Speaker Control
:SYSTem:SPEaker[:STATe] OFF|ON|0|1
:SYSTem:SPEaker[:STATe]?
Turns the internal speaker on or off.
Factory Preset
and *RST:
Off
Front Panel
Access:
Det/Demod, Speaker On Off
5-134
Chapter 5