Language Reference
FORMat Subsystem
This command specifies the formats used for trace data during data
NOTE
transfer across any remote port.
For corrected trace data (:TRACe[:DATA] with parameter
<trace_name>), REAL and ASCii formats will provide trace data in
the current amplitude units. INTeger format will provide trace data in
mdBm. The fastest mode is INTeger,32.
For uncorrected trace data (:TRACe[:DATA] with parameter
RAWTRACE), UINTeger and INTeger formats apply to RAWTRACE
queries, and return uncorrected ADC values. The fastest mode is
UINTeger,16.
For state data, the format cannot be changed. It is always in a machine
readable format only (machine units).
Table 5-1
ASCii
INT,32 (fastest)
REAL,32
REAL,64
Table 5-2
INT,32
UINT,16 (fastest)
5-52
Corrected Trace Data Types
:TRACe:DATA?<trace_name>
Data Type
Amplitude Units
Internal Units
Amplitude Units
Amplitude Units
Uncorrected Trace Data Types
:TRACe:DATA? RAWTRACE
Data Type
Uncorrected ADC
Values
Uncorrected ADC
Values
ASCii - Amplitude values are in ASCII, in amplitude units,
separated by commas.
INTeger, 32 - Binary 32-bit integer values in internal units (mdBm),
in a definite length block.
REAL, 32 (or 64) - Binary 32-bit, or 64-bit, real values in amplitude
units), in a definite length block.
Result
Result
Chapter 5