Select Tv Signal Path; Select Tv Standard - Agilent Technologies ESA-E Series Programmer's Manual

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Language Reference
TRIGger Subsystem
Remarks:
Normal baseband video has sync pulses on the bottom
of the waveform (use the
negative voltage). However, when the analyzer is used
to demodulate an NTSC or PAL TV RF carrier, the
detected video waveform is "upside down" with sync
pulses on the top of the waveform (use the
parameter for more positive voltage). When the
analyzer is used to demodulate a SECAM TV RF
carrier, the detected video waveform has normal
polarity (use the
Front Panel
Access:
Trig, TV Trig Setup, Sync Pos Neg

Select TV Signal Path

:TRIGger[:SEQuence]:TV:SOURce SANalyzer|EXTernal
:TRIGger[:SEQuence]:TV:SOURce?
This command is used to select between the internal spectrum analyzer
signal path (the detected video is fed to the TV trigger), or the EXT
VIDEO IN connector on the analyzer rear panel.
Factory Preset
and *RST:
SANalyzer
Front Panel
Access:
Trig, TV Trig Setup, TV Source, SA
Trig, TV Trig Setup, TV Source, EXT Video In

Select TV Standard

:TRIGger[:SEQuence]:TV:STANdard MNTSc|JNTSc|MPAL|BPAL|
NPAL|CPAL|LSEC
:TRIGger[:SEQuence]:TV:STANdard?
This command allows you to choose one of the various TV standards.
Factory Preset
and *RST:
NTSC-M
Example:
:TRIG:SEQ:TV:STAN MNTS
Remarks:
Once this function is defined, the selected type is
persistent. Persistent means that it retains the setting
previously selected, even through a power cycle. The
setting will change with "Load State."
5-148
parameter for more
NEG
parameter).
NEG
POS
Chapter 5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents